[Generation functions]
Circuit method : Bipolar sink and source
Constant Voltage: 2.5 V: 0 to ±2.5000 V (±0.03 % of setting ±300 μV, 100 μV resolution)
25 V: 0 to ±25.000 V (±0.03 % of setting ±3 mV, 1 mV resolution)
Constant Current :25 mA: 0 to ±25.000 mA (±0.03 % of setting ±3 μA, 1 μA resolution)
Thermoelectric power generation :K: at TC: 0 °C, -174.0 to 1372.0 °C (±0.05 % of setting ±0.5 °C, 0.1 °C resolution), Other types: E, J, T, R, S, B, N selectable
Thermoelectric power generation :K: at TC: RJ, -174.0 to 1372.0 °C (±0.05 % of setting ±1.0 °C, 0.1 °C resolution), Other types: E, J, T, R, S, B, N selectable
Standard resistance (Rs) :100 Ω (±0.2 Ω)
Automatic generation :Number of memory steps: 20 Interval time: 1 to 99 sec (at CV, CC, TC mode)
[Measurement functions]
Voltage :2.5 V: 0 to ±2.8000 V (±0.03 % rdg. ±300 μV, 100 μV resolution, 1 MΩ input resistance)
25 V: 0 to ±28.000 V (±0.03 % rdg. ±3 mV, 1 mV resolution, 1 MΩ input resistance)
Current :25 mA: 0 to ±28.000 mA (±0.03 % rdg. ±3 μA, 1 μA resolution, 25 Ω input resistance)
Temperature :-25.0 to ±80.0 °C (±0.5 °C at 23 ±5 °C, 0.1 °C resolution, use with the RJ sensor 9184)
Sampling rate :Approx. 1.67 times/sec
Additional functions: Zero adjustment, Overflow display, USB communication, Monitor
Power supply :AC ADAPTER 9445-02/03 (100 to 240 V AC 50/60 Hz, 9 VA), Ni-MH battery HR6 × 4, 6 VA, (fully charged 2500 mAh Ni-MH batteries: 170 minutes continuous use), or LR6 (AA) alkaline battery × 4, 6 VA
Dimensions, mass :104 mm (4.09 in) W × 180 mm (7.09 in) H × 58 mm (2.28 in) D, 660 g (23.3 oz) (including LR6 × 4 batteries)
Supplied accessories :INPUT CORD 9168 × 1, TEST LEAD L9170-10 × 1, Fuse × 1, LR6 (AA) alkaline battery × 4, Operation manual × 1